Method: reserve reagent volume before adding lysate
Addgene’s Western blot protocol specifies equivalent total protein amounts and dilution of loading buffer to final 1X. It also notes that suitable protein loading depends on the sample and target. Subramaniam Lab’s Laemmli instructions support both 2X and 4X stocks diluted to final 1X.
Thermo Fisher’s sample-preparation protocol includes loading buffer and optional separate reducer within the final mixture volume. The worksheet extends that accounting across different lysate concentrations; it does not establish compatibility between arbitrary reagents.
Symbols: M = protein mass per lane; V = final lane volume; C = lysate concentration; B = buffer stock factor; n = replicate lanes; e = excess percentage.
Loading buffer: L = V ÷ B
Separate reducer: R = V ÷ 10 when selected; otherwise zero
Available for lysate and diluent: A = V − L − R
Lysate: S = M ÷ C; diluent: D = A − S
Common maximum: minimum of C × A across all samples
Preparation multiplier: F = n × (1 + e ÷ 100)
Prepare each component × F. Load only V per lane from a homogeneous preparation.
The common maximum and preparation multiplier are algebraic extensions of the cited volume accounting, not experimentally determined loading recommendations. Exact zero diluent is valid; negative diluent blocks the entire preparation sheet.
Keep preparation and loading separate
- More mixture does not repair a crowded lane. Replicate lanes and excess scale every component together. They do not change per-lane mass, volume or feasibility.
- Do not count reducer twice. “None added separately” can still describe a reducing preparation. The Subramaniam Lab formulation, for example, already includes a reducing component.
- Use appropriate experimental conditions. This worksheet assumes corrected lysate concentrations, additive volumes, homogeneous mixtures, compatible reagents and enough lysate in each tube. Choose your own mass, diluent and gel-compatible volume. It does not select heating conditions or track tube inventory.
- Distinguish a rounded display from a usable bound. Exact decimal fractions determine feasibility. Masses and volumes normally display to two decimal places using half-up rounding. Any suggested two-decimal replacement target is rounded down, so it cannot exceed the exact common maximum. If no positive two-decimal target fits, the worksheet states that explicitly instead of recommending zero.
- Check pipetting resolution. Positive preparation components that would display as zero are flagged. Display precision is not a statement of equipment accuracy.
Research-use arithmetic, not a clinical diagnostic tool or a complete experimental protocol. Equal total protein is not evidence of assay linearity and does not replace loading controls. Numeric inputs accept decimal-point notation without scientific notation, separators or unit suffixes; each is limited to 120 characters.